Label update on handling of the finger grip for HEALAFLOW products
Healthcare professional
Medical devices
12 August 2016
This article has been migrated from an earlier version of the site and may display formatting inconsistencies.
Swissmed Pte Ltd would like to notify healthcare professionals of an update to the Instruction For Use (IFU) on handling of the finger grip for HEALAFLOW products following a recent incident that occurred during HEALAFLOW injection. During the injection, the finger grip of the syringe dissociated from the body of the syringe and led to an unexpected movement of the syringe and cannula into the patient’s eye. Healthcare professionals are advised to be aware of the proper handling of the syringe to avoid the dissociation of the finger grip during injection. Information on handling of the syringe will be updated in the IFU of HEALAFLOW. Please refer to the letter for details.
Only abstracts of the Dear Healthcare Professional Letters (DHCPL) are provided on this website. Details of these letters can be found in the MOH Alert System, which is accessible via the Health Professionals Portal (HPP), a one-stop portal for all healthcare professionals in Singapore. Dentists, medical doctors and pharmacists may access HPP at https://hpp.moh.gov.sg and follow the procedures below to access the MOH Alert system.
Step 1: Select your profession under ‘For Healthcare Professionals’.
Step 2: Click on ‘Useful links’.
Step 3: Click on ‘MOH Alert’ and log in via Singpass.
Step 4: Click on ‘Drug Alerts’ or ‘Med Device Alerts’ to view the DHCPLs.