Technical programming information for Cardiac Resynchronization Therapy pacemakers (VALITUDE™, VISIONIST™) and defibrillators (CHARISMA™ , MOMENTUM™, RESONATE™, VIGILANT™, AUTOGEN™, DYNAGEN™, INOGEN™, ORIGEN™) - Left Ventricular Offset Interaction
Healthcare professional
Medical devices
10 January 2018
This article has been migrated from an earlier version of the site and may display formatting inconsistencies.
Boston Scientific Asia Pacific Pte Ltd would like to inform healthcare professionals of the unintended asynchronous biventricular (BiV) pacing behaviour in Cardiac Resynchronization Therapy pacemakers (VALITUDE™, VISIONIST™) and defibrillators (CHARISMA™ , MOMENTUM™, RESONATE™, VIGILANT™, AUTOGEN™, DYNAGEN™, INOGEN™, ORIGEN™). This letter outlines the important reprogramming information to prevent this behaviour when tracking elevated atrial intrinsic rhythms in affected devices. Repeated detection of this unintended asynchronous BiV pacing behaviour may result in the implanted device reverting to a permanent Safety Mode (Safety Core™) status thus requiring early replacement. Healthcare professionals are advised to follow the enclosed programming recommendations to eliminate the risks associated with early device replacement. Please refer to the letter for details.
Only abstracts of the Dear Healthcare Professional Letters (DHCPL) are provided on this website. Details of these letters can be found in the MOH Alert System, which is accessible via the Health Professionals Portal (HPP), a one-stop portal for all healthcare professionals in Singapore. Dentists, medical doctors and pharmacists may access HPP at https://hpp.moh.gov.sg and follow the procedures below to access the MOH Alert system.
Step 1: Select your profession under ‘For Healthcare Professionals’.
Step 2: Click on ‘Useful links’.
Step 3: Click on ‘MOH Alert’ and log in via Singpass.
Step 4: Click on ‘Drug Alerts’ or ‘Med Device Alerts’ to view the DHCPLs.